Gas Sensor
Manning EC-P2
Instruction and Installation Manual
07/09
Release G Draft
Honeywell Confidential & Proprietary
This work contains valuable, confidential, and proprietary information. Disclosure, use
or reproduction outside of Honeywell Inc. is prohibited except as authorized in writing.
This unpublished work is protected by the laws of the United States and other countries.
Manning EC-P2 Sensor 19546 ECP2 07/09 REVG Copyright © 2009 Honeywell Analytics. All Rights Reserved.
1
About This Document
World Wide Web
The following Honeywell web sites may be of interest.
Honeywell Organization
WWW Address (URL)
Corporate
Honeywell Analytics
Manning Gas Detection
Telephone
Contact us by telephone at the numbers listed below.
Organization
Phone Number
United States
Honeywell Analytics Inc.
1-800-538-0363
1-913-712-5576
1-913-712-5580 Fax
Canada
Honeywell Analytics Inc.
Honeywell PACE
1-888-749-8878
Europe
+44 (0)1202 676161
+82 (0)2 2025 0307
+971 4 3458 338
Asia Pacific
Middle East
Honeywell Asia Pacific Inc.
Honeywell Analytics Inc.
Sales Information
Contact us at [email protected]
Symbol Definitions
The following table lists those symbols used in this document to denote certain conditions.
Symbol
Definition
ATTENTION: Identifies information that requires special consideration.
TIP: Identifies advise or hints for the user, often in terms of performing a task.
REFERENCE-EXTERNAL: Identifies an additional source of information outside of
this bookset.
REFERENCE-INTERNAL: Identifies an additional source of information within this
bookset.
Manning EC-P2 Sensor 19546 ECP2 07/09 REVG Copyright © 2009 Honeywell Analytics. All Rights Reserved.
3
Contents
Serial number:
Section
Title
Page
1
2
3
4
5
6
System Description
Operation
Maintenance
Programming and Advanced Features
Appendices
Limited Warranty
Introduction
This manual has been prepared to help in the use of the Manning EC-P2 Portable Gas Detector.
ATTENTION: This manual must be carefully followed by all individuals who have or will have the
responsibility for using or servicing the sensor. Warranties made by Honeywell Analytics with respect
to this equipment will be voided if the equipment is not used and serviced in accordance with the instructions in this
manual. If in doubt about a procedure, please contact Honeywell Analytics before proceeding.
Manning EC-P2 Sensor 19546 ECP2 07/09 REVG Copyright © 2009 Honeywell Analytics. All Rights Reserved.
4
1 System Description
Figure 1: Basic Parts of the Manning EC-P2 Sensor
The Manning EC-P2 is a portable, battery operated
instrument for the measurement of gas concentrations
in ambient air.
Outlet Tube
Sampling Wand
Assembly
The Manning EC-P2 can be converted from one
target gas to another by simply plugging in a different
SMART-CELL for the desired target gas (takes about
30 seconds). This allows one instrument to detect a
number of different gases. All calibration and alarm
setpoint data is stored on the individual SMART-CELL
so reconfiguration is not necessary when changing
target gases. Extra SMART-CELLs should be kept on
the optional sensor keeper when not in use to keep
them “warmed up” and ready for immediate use.
The Manning EC-P2 utilizes two different batteries to
insure that you’ll never be without power when you
need it. The primary power source is a “C” size
alkaline battery housed in the handle. It will run the
unit approximately 48 hours. When this battery is
removed or dead, the unit will switch to its standby
power source, an internal rechargeable NiCad
battery, which will run the unit approximately six
hours. When not in use, the unit should be plugged
into the charger to keep this standby battery charged.
If desired, the unit can be run continuously when
plugged into the charger.
Data Logger Port
Standby Battery
Recharge Receptacle
Primary Battery
The Manning EC-P2 also has advanced capabilities
such as data logging to an Excel spreadsheet, 0 to
1.0 Volt DC analog output capability, and a sampling
mode for taking a single reading in a controlled,
repeatable manner.
Manning EC-P2 Sensor 19546 ECP2 07/09 REVG Copyright © 2009 Honeywell Analytics. All Rights Reserved.
5
1 System Description
System Specifications
continued
Operating Ambient Temperature Range: 0o to
+120o F
Primary Power: Alkaline C cell battery, 48 hours
operation. Note: An audible alarm single beep every
60 seconds indicates that either the Alkaline C cell
battery is low or the internal rechargeable NiCad
battery is low.
Storage Temperature: Recommend storing on the
charger in an air conditioned office
Weight: 3 lbs. (shipping weight), 1.8 lbs. (detector
only)
Standby Power: Internal rechargeable NiCad
battery, 6 hours operation. Note: An audible alarm
single beep every 60 seconds indicates that either
the Alkaline C cell battery is low or the internal
rechargeable NiCad battery is low.
Dimensions: Approx 3.5" x 5" x 8.5"
Outputs: RS-232 output of stored gas values.
Analog output 0 to 1.0 VDC (requires optional analog
output cable)
Charger: 120 VAC plug-in charger for internal
Memory: 12,000 data points, storage interval
between points programmable at 1, 5, 10, or 15
minutes. Typical capacity: 8 days at 1 minute storage
interval
standby battery
Display: Backlit LCD
Pump: Miniature internal diaphragm type,
flowrate 0.5 SCFH
Alarms: Two concentration alarms, zero drift
caution, Low flow and Low battery alarms. Alarms
displayed on LCD and indicated by audible beeper.
Standard Accessories: Storage case, 10" extension
wand, battery charger, filter material, datalogging
software, RS-232 output cable (DB9 connector)
Optional Accessories: Analog output cable. Sensor
Keeper for storage of additional SMART-CELLs
Sensor Type: Target gas selective electrochemical
SMART-CELL
Accuracy: Sensor dependant but generally ± 5% of
reading
Repeatability: ± 1% full scale
Manning EC-P2 Sensor 19546 ECP2 07/09 REVG Copyright © 2009 Honeywell Analytics. All Rights Reserved.
6
2 Operation
A Operating the Sensor
4)
Replace the manifold and hand tighten the
thumbscrews.
The Manning EC-P2 is shipped ready for use. To
power up the instrument, push the PWR button. The
power up sequence displays first the self check, the
sensor full scale value, the downscale caution level,
warning concentration setpoint, alarm concentration
setpoint, output ranges, software version, then the
normal operating screen.
5)
Verify proper flow through the sampling wand
using the flowmeter provided.
Extra SMART-CELLS should be kept on the
optional sensor keeper when not in use to
keep them “warmed up” and ready for immediate
use. The sensor keeper is powered by an Alkaline
“C” size battery which should be replaced every 6
months.
The unit will then display the instantaneous gas
concentration and is ready for use.
The normal operating screen displays the information
as shown in Figure 2.
available SMART-CELLs. Contact Honeywell Analytics
if the desired target gas is not listed, as new cells may
be available.
The Manning EC-P2 is designed to avoid accidental
shut-off of the power. To turn the unit off, the PWR
button must be held down for approximately three
seconds. Release the button after the “Power Down”
message appears.
Figure 2: Normal Operating Screen
BATTERY IN USE
P = Primary (“C” cell)
CURRENT GAS CONCENTRATION
(Blinking indicates sensor over range)
S = Standby (Internal NiCad)
Flashing indicates low battery
TARGET GAS
SMART-CELL dependent
B Changing the Target Gas
using SMART-CELL
Technology
PPM
P
A
The Manning EC-P2 can easily be changed to sense
another target gas by installing the SMART-CELL for
the desired target gas. Each SMART-CELL contains
all the calibration and alarm setpoint information on
its internal memory chip, so recalibration or
reprogramming is not necessary when changing the
Manning EC-P2 from one gas to another. This
procedure can be performed with the unit powered
on or off (see Figure 3).
NH
3
0
OFF
MENU SMPL
Battery Light
Power
ADVANCED FEATURES ONLY
1)
Unscrew the two stainless steel thumbscrews
and remove the black inlet / outlet manifold lid
from the back of the unit.
Do not push for normal operation
ALARM INDICATION CODE
Blank = No alarm condition present
C = Caution—Cell has excessive zero downdrift
W = Warning—Gas concentration has exceeded warning level
A = Alarm—Gas concentration has exceeded alarm level
P = Pump trouble
2)
3)
To remove the existing SMART-CELL, grip the
sides of the cell and pull it straight out.
Install the desired SMART-CELL by gently
inserting it into the socket. Notice that the
SMART-CELL is keyed so it cannot be inserted
incorrectly.
Lg = Data logging in progress
Manning EC-P2 Sensor 19546 ECP2 07/09 REVG Copyright © 2009 Honeywell Analytics. All Rights Reserved.
7
2 Operation
continued
C Response Test
In addition to the alarm indication code in the left side
of the screen, an alarm message will flash inter-
mittently on the screen, the audible beeper will sound
and the ACK button will appear. Pushing the ACK
button will silence the beeper and stop the display
flashing. The Alarm Indication Code will remain on
the left side of the screen until the alarm condition
clears.
Prior to use, the unit should first be tested for proper
response. With the detector operating, the unit
should be exposed to a sample of the target gas.
The display should show an increasing concentra-
tion. If not, do not use it for field measurements.
D Sample Inlet Port
Alarm setpoints are stored on the individual SMART-
CELL and can be changed by the user as described
in Section 4.
The Manning EC-P2 is provided with a quick disconnect
flexible extension wand. It will work with or without the
wand. When using the extension wand, the internal
walls must be kept dry. Water on the walls of the tubing
can potentially absorb the target gas. To dry, allow the
unit to pump dry, ambient air for 15 minutes.
NOTE: An audible alarm single beep every
60 seconds indicates that either the Alkaline
C cell battery is low or the internal rechargeable
NiCad battery is low.
E Response/Recovery Time
H Display Resolution
Under normal conditions, the instrument will reach
90% of final value within two minutes. This is
dependent on concentration and temperature.
The Manning EC-P2 displays gas concentration with
a resolution that depends upon the range of the unit.
Full scale ranges of 0-4.00 or below will provide
resolution of 0.01. Ranges from 0-5.0 up to 0-49.9 will
provide resolution of 0.1. Full scale ranges above 50
will have a resolution of 1.
Recovery time for the sensor depends on duration
and concentration of exposure. Short exposures of
concentrations at the lower end of the target gas
range result in rapid recovery. Long exposures to
levels above the middle of the target gas range or
short exposures to levels exceeding the target gas
range can extend recovery times to hours. Repeated
exposures above the target gas range will reduce cell
life and should be avoided.
Figure 3: Manifold Assembly
Outlet tube
barb fitting
Thumbscrews
(loosen to remove
manifold lid)
TEST UNIT PRIOR TO USE
Sensor
(SMART-CELL)
OUTLET
Filter Material
F Interference Gases
Sampling
wand
INLE
CAUTION
The Manning EC-P2 is generally quite specific to the
target gas, but depending on the target gas, other
gases may cause a reading. See Appendix B for a
listing of potential interference gases for the various
SMART-CELLs.
DO NOT BLOCK INLET
OR OUTLIET PORTS
Inlet fitting
Sensor
alignment
groove
Manifold lid
G Alarm Functions
The Manning EC-P2 provides both visible and
audible gas concentration alarms and system
alarms. An alarm condition will be displayed on the
normal operating screen as described on page 8.
Sensor
alignment
pin
Manning EC-P2 Sensor 19546 ECP2 07/09 REVG Copyright © 2009 Honeywell Analytics. All Rights Reserved.
8
3 Maintenance
A Periodic Maintenance
C Calibration
It is essential that the test and calibration schedule
be adhered to. Honeywell Analytics recommends the
following maintenance schedule:
Calibration is recommended every six months in
normal use. Each SMART-CELL stores its calibration
and alarm setpoint information on its own memory
chip. When the SMART-CELL is plugged into the
Manning EC-P2, this information is used by the
sensor. The SMART-CELL can be field calibrated by
the user or returned to Honeywell Analytics for
factory calibration.
•
Response test prior to each use. Expose
sensor to a sample of the target gas to verify
proper sensor response and alarm functions.
•
Calibration should be performed with certified
calibration gas every six months, and all tests
must be logged.
The calibration procedure requires the use of a
bypass tee to allow the gas to be drawn into the flow
cell without creating variable pressure or dilution
problems. Calibration gas cannot be fed into the flow
cell under pressure.
•
•
Check filter material every six months. Replace
if dirty or moist.
If the optional sensor keeper is used for extra
SMART-CELLs, change the sensor keeper
battery once every 6 months. Use only alkaline
“C” size batteries.
When applying calibration gas use the bypass tee as
shown in Figure 4 on page 11. The flowrate of
calibration gas should be high enough that at least
0.5 SCFH is flowing out of the calibration tee. This
can be measured with the supplied flowmeter and
insures that air is not being drawn into the unit and
diluting the calibration gas.
B Sensor Life
The electrochemical SMART CELL is extremely
reliable, but the following can cause the cell
chemicals to become depleted:
The unit may be zeroed without spanning it; however,
never span the unit without first zeroing it.
•
•
•
A period of time
The following procedure will calibrate the SMART-
CELL.
Exposure to high temperatures
Exposure to high concentrations of
the target gas
To Enter ZERO / SPAN Mode
1)
Push power button to turn unit on. Wait for
power up sequence to finish (30 seconds).
•
Exposure to high moisture
When the cell becomes depleted, the unit will give no
indication of failure other than that the sensor will not
respond. For this reason it is absolutely essential that
these units be exercised with a gas sample on a
regular basis.
2)
Push and hold the unnamed button (second
from left) first, then Push and hold the Menu
button. Hold both buttons until the screen goes
blank (2 seconds). Release both buttons and
the unit is now in Programming Mode.
Typical sensor life under normal use is two years or
more. When the SMART-CELL will no longer calibrate
properly, simply purchase a new SMART-CELL and
replace.
3)
Push the SENS button and the unit is now in the
zero/span mode.
Manning EC-P2 Sensor 19546 ECP2 07/09 REVG Copyright © 2009 Honeywell Analytics. All Rights Reserved.
9
3 Maintenance
continued
Zero Calibration
Unit does not respond when keys are pressed:
Press and hold all four buttons for five seconds. This
will reset the Manning EC-P2. Turn the unit back on
by pressing the PWR button.
4)
Be sure the unit is in clean air or apply zero
air for two minutes.
5)
Push ZERO button.
High reading won’t clear: When the sensor is
exposed to very high gas concentrations, it takes a
relatively long time for the sensor to clear. Leave the
unit running in a clean environment for a few hours to
clear the sensor. If it doesn’t clear after six hours, the
sensor may have been damaged. Contact Honeywell
Analytics for a replacement SMART-CELL.
6)
Push SAVE button to save the new zero
calibration.
Span Calibration
7)
8)
9)
Push SPAN button.
Apply span gas for two minutes.
Unit is in alarm with gas reading of 0 ppm: Check
the alarm setpoints and alarm functions as described
downscale alarm, and that the setpoint is above zero.
Push INC button (increase) or DEC button
(decrease) until the display matches the span
gas concentration.
10) Push SAVE button to store new span calibration.
Figure 4: Calibration of the Manning EC-P2 Sensor
using a Bypass Tee
To Exit ZERO / SPAN Mode
11) Push DONE button twice slowly to get back into
normal operation mode.
Vent through
flowmeter
Bypass tee
D Troubleshooting
The detector does not start: Check to see that a
fresh battery is installed. Both the primary and
standby batteries must be discharged for this
problem to occur. Plug the charger into the Manning
EC-P2 to charge the standby battery. Wait 10
minutes and retry.
TEST UNIT PRIOR TO USE
OUTLET
INLE
1.0
.8
CAUTION
DO NOT BLOCK INLET
OR OUTLIET PORTS
.6
Calibration gas
flowmeter
The detector does not respond to the target gas:
Verify the proper flow rate using the supplied
flowmeter. Perform calibration. If sensor won’t
calibrate, the SMART-CELL may be depleted.
Contact Honeywell Analytics for a replacement
SMART-CELL.
.4
.2
LPM AIR
Calibration gas bottle
(use only Certified
Calibration Gas)
Pump alarm: This will occur if the pump motor stops
or if an internal pressure sensor detects blockage of
the inlet. Check that the internal filter is not clogged
or wet and that the inlet tube is clear.
External flowmeter indicates no flow: Verify that
pump is running. Check that the manifold screws are
tight and the extension wand connection is tight.
Manning EC-P2 Sensor 19546 ECP2 07/09 REVG Copyright © 2009 Honeywell Analytics. All Rights Reserved.
10
4 Programming and Advanced Features
A Alarm Setpoints, Functions
gas concentration (DOWN arrow), or be disabled
(off). The following procedure will set the Warning
function to alarm on decreasing concentration
(Useful for Oxygen Depletion warning).
The Manning EC-P2 has three adjustable alarm
setpoints. These setpoints are stored on each
SMART-CELL. The Caution setpoint is to warn of
excessive negative zero drift. The WARNING and
ALARM setpoints are to warn of potentially
To Enter CAUT / WARN / ALRM Mode
dangerous target gas concentrations. The following
procedure will change the WARNING setpoint.
1)
Push power button to turn unit on. Wait for
power up sequence to finish (30 seconds).
2)
Push and hold the unnamed button first, then
Push and hold the Menu button. Hold both
buttons until the screen goes blank (two
seconds). Release both buttons and the unit is
now in Programming Mode.
To Enter CAUT / WARN / ALRM Mode
1)
Push power button to turn unit on. Wait for
power up sequence to finish (30 seconds).
2)
Push and hold the unnamed button first, then
Push and hold the Menu button. Hold both
buttons until the screen goes blank (two
seconds). Release both buttons and the unit is
now in Programming Mode.
3)
Push the ALRMS button and the unit is now in
the CAUT / WARN / ALRM mode.
Warning Function
4)
5)
6)
Push WARN button.
3)
Push the ALRM button and the unit is now in the
CAUT / WARN / ALRM mode.
Push FUNC (function) button.
Warning Setpoint
Push SELECT button while viewing the arrow in
the lower left hand corner of the screen to
toggle through the following three possibilities:
4)
5)
6)
Push WARN button.
Push S.P. button (setpoint).
•
•
UP arrow — Warn on increasing concen-
tration (gas goes higher than setpoint).
DOWN arrow — Warn on decreasing
concentration (gas goes lower than
setpoint).
Push INC button (increase) or DEC button
(decrease) until the display matches the desired
warning setpoint. Notice the minus sign above
the arrow if the setpoint is below zero.
•
Blank — warning function turned OFF.
7)
8)
Push SAVE button to store new warning
setpoint.
7)
8)
Push SAVE button when the DOWN arrow is
displayed to store new warning function.
Push DONE button to return to Alarms mode.
Push DONE button to return to Alarms mode.
To Exit CAUT / WARN / ALRM Mode
To Exit CAUT / WARN / ALRM Mode
9)
Push DONE button twice slowly to get back into
normal operation mode.
9)
Push DONE button two times slowly to get back
into normal operation mode.
The Caution and Alarm setpoints are pro-
grammed in the same manner, except for step
four above in which you would push either the
CAUT or ALRM button. The function of each alarm
can be selected to alarm on increasing gas
concentration (UP arrow), alarm on decreasing
The Caution and Alarm functions are pro-
grammed in the same manner, except for step
four above in which you would push either the
CAUT or ALRM button.
Manning EC-P2 Sensor 19546 ECP2 07/09 REVG Copyright © 2009 Honeywell Analytics. All Rights Reserved.
11
4 Programming and Advanced Features
continued
B Display Variables
C Sample Mode
The Manning EC-P2 has two adjustable display
variables. These variables control signal conditioning
of the LCD display. Variables are stored on each
SMART-CELL. The AVG (Average) variable
determines the number of one second samples to
average to calculate the displayed value. The lower
the averaging value, the faster the display will
change and the more “wiggle” will be displayed. The
BLANK (zero blanking) variable controls the lowest
reading that will be displayed. This is to eliminate the
misleading display of signals within the zero noise of
the sensor. The following procedure will change the
AVG variable.
The Sample Mode feature allows the user to take a
measurement using a predefined repeatable method
to eliminate user “interpretation” of a gas concentra-
tion that is changing. This is similar to a digital scale,
for example, so the exact same reading will be taken
regardless of who is taking it.
The measuring sequence used in Sample Mode is as
follows:
•
Draw a Sample for a user programmed
sampling time to allow unit to stabilize to the
existing gas concentration.
•
•
Measure the gas concentration and average it
for a user programmed measuring time.
To Enter AVG / BLANK Mode
Clear the gas concentration to below a user
programmed recovery value before allowing
another cycle to start. (Unit should be moved to
fresh air during the clear time.)
1)
Push power button to turn unit on. Wait for
power up sequence to finish (30 seconds).
2)
Push and hold the unnamed button first, then
Push and hold the Menu button. Hold both
buttons until the screen goes blank (two
seconds). Release both buttons and the unit is
now in Programming Mode.
Taking a sample:
To take a sample, locate the detector at the desired
sample location, and push the SMPL (sample) button
on the normal operation screen. The unit will count
down the Sample time, followed by the Measure time,
at the end of which the average value will be
displayed.
3)
4)
Push the MORE button.
Push the DISP (display) button and the unit is
now in the AVG / BLANK mode.
AVG Variable
Note the displayed sample value. Pushing the
CLEAR button will return the unit to the normal
operation screen once the gas concentration is
below the predefined recovery value. The unit is now
ready to take another sample if desired.
5)
6)
Push AVG button.
Push INC button (increase) or DEC button
(decrease) until the display matches the desired
value.
7)
Push SAVE button to store new AVG variable.
To Exit AVG / BLANK Mode
8)
Push DONE button twice slowly to get back into
normal operation mode.
The Blanking variable is programmed in the
same manner, except for step 5 above in which
you would push the BLANK button.
Manning EC-P2 Sensor 19546 ECP2 07/09 REVG Copyright © 2009 Honeywell Analytics. All Rights Reserved.
12
4 Programming and Advanced Features
continued
Programming sample mode variables:
EC-P2 software installation:
The sample mode variables are Sample, Measure
and Clear as described above. The following
procedure will change the Sample variable.
The Manning EC-P2 is supplied with a CD containing the
“Manning EC-P2” program. It will run on Windows 95,
Windows 98, or Windows 2000. To install the program,
place the disk in your CD drive. The CD should
automatically begin the installation process. Follow the
prompts as they appear on your screen. Accepting the
default directory locations is recommended for most
installations. If the CD does not automatically start,
click on “My Computer” and then your CD drive.
Double-click on the file named “Setup.exe.”
To Enter SAMP / MEAS / CLEAR Mode
1)
Push power button to turn unit on. Wait for
power up sequence to finish (30 seconds).
2)
Push and hold the unnamed button (second
from left) first, then Push and hold the Menu
button. Hold both buttons until the screen goes
blank (two seconds). Release both buttons and
the unit is now in Programming Mode.
Prior to your first datalogging session it is recom-
mended that you connect the Manning EC-P2 to your
computer with the data logging program running.
The computer can then set the time in the EC-P2 to
match its own clock.
3)
4)
Push the MORE button.
Push the SMPL (sample) button and the unit is
now in the SAMP / MEAS / CLEAR mode.
Data Logging with the Manning EC-P2:
To Start the Data Logging Session
SAMP Variable
5)
6)
Push SAMP (sample) button.
1)
Push power button to turn unit on. Wait for
power up sequence to finish (30 seconds).
Push INC button (increase) or DEC button
(decrease) until the display matches the desired
value. The display is in minutes and seconds.
2)
3)
Push the MENU button.
Push the LOG button and the unit is now in the
CLEAR / INT / START mode.
7)
Push SAVE button to store new SAMP variable.
To Exit SAMP / MEAS / CLEAR Mode
4)
If the CLEAR button is available, that means
there is data currently stored in the unit. If
desired, you can erase the data already in the
unit by pushing the CLEAR button. (If you don’t
push CLEAR, each data logging session can be
downloaded as a separate file).
8)
Push DONE button twice slowly to get back into
normal operation mode.
The Measure and Clear variables are pro-
grammed in the same manner, except for step
five above in which you would push the MEAS
or CLEAR button.
5)
6)
Push INT button (interval) to cycle through the
choices for how often to log a data point. Stop
when the desired interval is displayed. Notice
how much time can be covered with the
remaining free memory.
D Data Logging
The Manning EC-P2 can log periodic gas concentra-
tion readings to be downloaded to a file which can be
read by Microsoft Excel or other popular spreadsheet
applications. The EC-P2 will store instantaneous gas
values every one, five, 10, or 15 minutes. Multiple
sessions can be logged and then downloaded as
separate files.
Push and hold the START button until the screen
goes blank (two seconds) to begin a new data
logging session. The unit will now display the gas
concentration. The Lg symbol in the left side of the
display indicates that the unit is logging data. If
you watch this symbol carefully, you will see it
change to an “s” each time a data point is stored.
Manning EC-P2 Sensor 19546 ECP2 07/09 REVG Copyright © 2009 Honeywell Analytics. All Rights Reserved.
13
4 Programming and Advanced Features
continued
To Stop the Data Logging Session
E Analog Output
7)
8)
9)
Push the MENU button.
Push the LOG button.
The Manning EC-P2 has a 0 to 1.0 volt DC analog
output capability when used with the optional analog
output cable. With the analog output cable plugged into
the communication port of the EC-P2, the user can
measure the voltage between the two wires and
correlate it to gas concentration. The voltage output will
be linear between zero and the full scale range of the
SMART-CELL as programmed below. Each SMART-
CELL has a wide range of choices of full scale values.
Push and hold the STOP button until the screen
changes to CLEAR / INT / START / MODE (two
seconds).
10) Push the DONE button to get back into normal
operation mode.
Working with the “EC-P2” program:
The full scale range as programmed below
only affects the analog output and the
The “EC-P2” program downloads the logging files
from the Manning EC-P2. It does not display the
data. The files it downloads can be opened by or
imported into Excel, Lotus 123 or other programs for
graphing and analysis.
data logging full-scale output. The display will
always cover the entire operating range of the
SMART-CELL. Changing the range can affect the
display resolution for some SMART-CELLs.
Turn on the Manning EC-P2 and connect it to your
PC using the RS-232 cable supplied. From Windows
START menu, select “PROGRAMS,” then “Manning
EC-P2,” and then “EC-P2”. As soon as you start the
“EC-P2” program, your computer will try to establish
a connection to the Manning EC-P2. If the Manning
EC-P2 is currently logging data, the “EC-P2” program
will automatically stop the logging function.
Changing the full scale range on EC-P2:
To Enter RANGE Mode
1)
Push power button to turn unit on. Wait for
power up sequence to finish (30 seconds).
2)
Push and hold the unnamed button (second
from left) first, then Push and hold the MENU
button. Hold both buttons until the screen goes
blank (two seconds). Release both buttons and
the unit is now in Programming Mode.
The program contains two tabs, one marked
“Download” and one marked “Configure”. Review the
settings on the “Configure” tab, making changes as
necessary.
3)
4)
5)
Push the MORE button.
From the “Download” tab, review the Link Status
block in the lower right hand corner. It should
indicate “Link OK”. Link Trouble can be caused by
selection of the wrong Com port in the Configure
Tab, improperly connecting the RS-232 cable, or not
turning on the Manning EC-P2.
Push the DISP (display) button.
Push the MORE button and the unit is now in the
RANGE mode.
Adjusting the RANGE
6)
7)
Push RANGE button.
When you are ready to transfer data from the
Manning EC-P2 to your PC, click on “get data” and
the files will be downloaded to your computer. Note
the file names in the destination window so you can
locate the files.
Push INC button (increase) or DEC button
(decrease) until the display matches the desired
full scale range.
8)
Push SAVE button to store new full scale range.
These files can now be opened with Excel (use .csv
filename), Lotus 123, or other user supplied
programs.
To Exit RANGE Mode
9)
Push DONE button twice slowly to get back into
normal operation mode.
Manning EC-P2 Sensor 19546 ECP2 07/09 REVG Copyright © 2009 Honeywell Analytics. All Rights Reserved.
14
5 Appendices
APPENDIX A: SMART-CELLs Available
GAS
RANGE
PART NUMBER
Alcohol
0-500 ppm
0-500 ppm
0-2000 ppm
0-5 ppm
EC-P2-SMART-CELL-OH-0/500ppm
EC-P2-SMART-CELL-NH3-0/500ppm
EC-P2-SMART-CELL-NH3-0/2000ppm
EC-P2-SMART-CELL-Br2-0/5ppm
EC-P2-SMART-CELL-CO-0/1000ppm
EC-P2-SMART-CELL-Cl2-0/5ppm
EC-P2-SMART-CELL-Cl2-0/200ppm
EC-P2-SMART-CELL-ClO2-0/5ppm
EC-P2-SMART-CELL-Fl2-0/5ppm
EC-P2-SMART-CELL-HCHO-0/200ppm
EC-P2-SMART-CELL-H2-0/2000ppm
EC-P2-SMART-CELL-HCl-0/200ppm
EC-P2-SMART-CELL-H2O2-0/200ppm
EC-P2-SMART-CELL-H2S-0/200ppm
EC-P2-SMART-CELL-NO-0/500ppm
EC-P2-SMART-CELL-NO2-0/200ppm
EC-P2-SMART-CELL-O2-0/25%
Ammonia
Ammonia
Bromine
Carbon Monoxide
Chlorine
0-1000 ppm
0-5 ppm
Chlorine
0-200 ppm
0-5 ppm
Chlorine Dioxide
Flourine
0-5 ppm
Formaldehyde
Hydrogen
0-200 ppm
0-2000 ppm
0-200ppm
0-200 ppm
0-200 ppm
0-500 ppm
0-200 ppm
0-25%
Hydrogen Chloride
Hydrogen Peroxide
Hydrogen Sulfide
Nitric Oxide
Nitrogen Dioxide
Oxygen
Ozone
0-5 ppm
EC-P2-SMART-CELL-O3-0/5ppm
EC-P2-SMART-CELL-O3-0/100ppm
EC-P2-SMART-CELL-SO2-0/500ppm
Ozone
0-100 ppm
0-500 ppm
Sulfur Dioxide
Manning EC-P2 Sensor 19546 ECP2 07/09 REVG Copyright © 2009 Honeywell Analytics. All Rights Reserved.
15
5 Appendices
continued
APPENDIX B: SMART-CELL Cross-Sensitivity Table (Interference Gases)
SENSOR AND TARGET GAS
Br2
Cl2
ClO2
O3
NH3
CO
H2
O2
NO
H2S
NO2
SO2
HCHO OH
Fl2
HCl
H2O2
Bromine
Chlorine
-----
1.0
1.0
-----
1.0
1.0
0.7
0.7
-0.1
-0.1
None None None None None 0.5
None None None None None 0.5
-0.1
-0.1
None None 1.0
None None 1.0
None None
None None
I
N
T
Chlorine dioxide
Ozone
1.0
1.5
1.0
1.5
-----
1.5
0.7
-----
-0.1
-0.1
None None None None None 0.5
None None None None None 0.65
-0.1
-0.1
None None 1.0
None None 1.5
None None
None None
E
R
F
E
R
E
N
C
E
Ammonia
Carbon monoxide
None None None None -----
None None None None 0.05
None None None None None None None None None None None None
-----
0.1
None None 0.002 0.001 None 0.5
0.5
None 0.005 0.005
Hydrogen
Oxygen
None None None None 0.02
0.1
-----
None 0.001 0.001 0.001 0.005 0.05
0.05
None 0.01 0.01
None None None None None None None -----
None None None None None 0.1 None None -----
None None None None -----
None None None None None 0.1
None None None None None None None None None
Nitrogen oxide
Hydrogen sulfide
0.4
0.001 0.04
0.8
2
0.8
2
None 1.5
1.5
4
-0.1
0.2
-0.1
0.2
-0.1
0.2
-0.07 0.3
0.15
-2.5
None
-0.1
3
Nitrogen dioxide
Sulfur dioxide
-----
-1
-0.08 0.1
0.1
0.4
0.2
-0.01 0.5
0.2
0.2
1
-0.01 -0.01 -0.01 -0.01 None None None None None 0.1
-----
0.4
Formaldehyde
Alcohol
None None None None 0.03
None None None None 0.03
0.1
0.1
0.1
0.1
None 0.05
None 0.05
0.01
0.01
None 0.02
None 0.02
-----
None -----
1.2
None None 0.1
None None 0.1
Flourine
1.0
1.0
1.0
0.7
-0.1
None None None None None 0.5
-0.1
None None -----
0.2 None -----
None None
0.1
Hydrochloric acidNone None None None None None None None None None -0.3
None 0.2
G
A
S
E
S
Hydrogen peroxide
Hydrogen cyanide
None None None None 0.03
-0.08 -0.08 -0.08 -0.05 None 0.1
None None None None None None None None None None None None None None None None None
0.1
0.1
None 0.05
0.01
None 0.02
None None None None -----
None None None None -0.07 0.15
0.1 0.1 -0.08 0.01 0.1
Hydroflouric acid
Hydride
None None None None 0.5
None None None
1
0.5
-2
2
2
2
None 1.5
2
Silica hydride
Carbon dioxide
None None None None 0.5
None None None None None None None None None None None None None None None None None
None None None None None None None None None None None None None None None None None
None None None
1
0.5
-2
2
2
2
None 1.5
2
Methane
Carbosulfonic acid
-0.04 -0.04 -0.04 -0.03 None None None None None 0.3
None None None None None 0.1 0.1 None None None None None
None None None None 0.01 None None None 0.001 0.005 0.001 0.05
-0.8
None
1
1
-0.04
None None None
None 0.01 0.02
1
1.3
Ethylene
Ethyl alcohol
1
2
1
2
Notes: 1) None means there is no interference.
2) A number indicates that when exposed to the interference gas, the sensor will react the specified amount for a given
concentration. For example, on the first line of the table above, if the NO2 sensor is exposed to 1 ppm Bromine (Br2),
the sensor will read 0.5 ppm. This interference is in addition to the target gas present.
For example, if an NH3 sensor is exposed to 1 ppm NH3 and 1 ppm Br2 at the same time, the reading will be 0.9 ppm
(1ppm NH3 + -0.1 ppm Br2 = 0.9 ppm reading).
Manning EC-P2 Sensor 19546 ECP2 07/09 REVG Copyright © 2009 Honeywell Analytics. All Rights Reserved.
16
6 Limited Warranty
1.
Limited Warranty
3.
Warranty Limitation and Exclusion
Honeywell Analytics, Inc. warrants to the
original purchaser and/or ultimate customer
(“Purchaser”) of Manning products (“Product”)
that if any part thereof proves to be defective in
material or workmanship within eighteen (18)
months of the date of shipment by Honeywell
Analytics or twelve (12) months from the date of
first use by the purchaser, whichever comes first,
such defective part will be repaired or replaced,
free of charge, at Honeywell Analytics’ discretion
if shipped prepaid to Honeywell Analytics at
405 Barclay Blvd., Lincolnshire, IL 60069,
in a package equal to or in the original container.
The Product will be returned freight prepaid and
repaired or replaced if it is determined by
Honeywell Analytics that the part failed due to
defective materials or workmanship. The repair
or replacement of any such defective part shall
be Honeywell Analytics’ sole and exclusive
responsibility and liability under this limited
warranty.
Honeywell Analytics will have no further
obligation under this limited warranty. All
warranty obligations of Honeywell Analytics are
extinguishable if the Product has been subject
to abuse, misuse, negligence, or accident or if
the Purchaser fails to perform any of the duties
set forth in this limited warranty or if the Product
has not been operated in accordance with
instructions, or if the Product serial number has
been removed or altered.
4.
Disclaimer of Unstated Warranties
THE WARRANTY PRINTED ABOVE IS THE
ONLY WARRANTY APPLICABLE TO THIS
PURCHASE. ALL OTHER WARRANTIES,
EXPRESS OR IMPLIED, INCLUDING, BUT NOT
LIMITED TO, THE IMPLIED WARRANTIES OF
MERCHANTABILITY OR FITNESS FOR A
PARTICULAR PURPOSE ARE HEREBY
DISCLAIMED.
5.
Limitation of Liability
IT IS UNDERSTOOD AND AGREED THAT
HONEYWELL ANALYTIC’S LIABILITY,
2.
Exclusions
A. If gas sensors are part of the Product, the gas
sensor is covered by a twelve (12) month
limited warranty of the manufacturer.
B. If gas sensors are covered by this limited
warranty, the gas sensor is subject to
inspection by Honeywell Analytics for
extended exposure to excessive gas con-
centrations if a claim by the Purchaser is
made under this limited warranty. Should
such inspection indicate that the gas sensor
has been expended rather than failed
prematurely, this limited warranty shall not
apply to the Product.
WHETHER IN CONTRACT, IN TORT, UNDER
ANY WARRANTY, IN NEGLIGENCE OR
OTHERWISE SHALL NOT EXCEED THE
AMOUNT OF THE PURCHASE PRICE PAID BY
THE PURCHASER FOR THE PRODUCT AND
UNDER NO CIRCUMSTANCES SHALL
HONEYWELL ANALYTICS BE LIABLE FOR
SPECIAL, INDIRECT, OR CONSEQUENTIAL
DAMAGES. THE PRICE STATED FOR THE
PRODUCT IS A CONSIDERATION LIMITING
HONEYWELL ANALYTICS’ LIABILITY. NO
ACTION, REGARDLESS OF FORM, ARISING
OUT OF THE TRANSACTIONS UNDER THIS
WARRANTY MAY BE BROUGHT BY THE
PURCHASER MORE THAN ONE YEAR AFTER
THE CAUSE OF ACTIONS HAS OCCURRED.
C. This limited warranty does not cover consum-
able items, such as batteries, or items subject
to wear or periodic replacement, including
lamps, fuses, valves, vanes, sensor elements,
cartridges, or filter elements.
Manning EC-P2 Sensor 19546 ECP2 07/09 REVG Copyright © 2009 Honeywell Analytics. All Rights Reserved.
17
|